SPECIFICATIONS |
MODEL | 7130 | 7132 | 7140 | 7142 |
AC WITHSTAND VOLTAGE |
Output Rating | 5KVAC/20mA |
Range | Resolution | Accuracy |
Output Voltage, KVAC | 0-5.00 | 0.01 | ±(2% of setting + 5V) |
Output Frequency | 50Hz/60Hz ±100ppm, User Selection |
Output Waveform | Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5 |
Output Regulation | ± (1% of output + 5V), From no load to full load |
SETTINGS |
Max-Limit AC Current, mA | 0 - 20.00 | 0.01 | ±(2% of setting + 2counts) |
Min-Limit AC Current, mA | 0 - 9.999 | 0.001 | ±(2% of setting + 2counts) |
Ramp Up Time, second | 0.1 - 999.9 | 0.1 | ±(0.1% + 0.05sec) |
Ramp Down Time, second | 0 - 999.9 | 0.1 |
Dwell Time, second | 0, 0.3 - 999.9 (0=continuous) | 0.1 |
Arc Detection | 0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity) |
DC WITHSTAND VOLTAGE (7140,7142 ONLY) |
Output Rating | 6KVDC/7500μA |
Output Voltage, KVDC | 0 - 6.00 | 0.01 | ±(2% of setting + 5V) |
Output Ripple | <5%(6KV/7500μA at Resistive Load) |
SETTINGS |
Max-Limit DC Current, μA | 0 - 7500 | 1 | ±(2% of setting + 2counts) |
Min-Limit DC Current, μA | 0 - 999.9 | 0.1 | ±(2% of setting + 2counts) |
Ramp Up Time, second | 0.1 - 999.9 | 0.1 | ±(0.1% + 0.05sec) |
Ramp Down Time, second | 0, 1.0 - 999.9 | 0.1 |
Dwell Time, second | 0, 0.4 - 999.9 (0=continuous) | 0.1 |
Arc Detection | 0, 1 - 9 ranges (0=OFF, 9 is the most sensitivity) |
Discharge Time | < 200 msec |
INSULATION RESISTANCE (7132,7142 ONLY) |
Output Rating | 1KVDC/9999MΩ |
Output Voltage, DCV | 30 - 1000 | 10 | ±(2% of setting + 5V) |
SETTINGS |
Max-Limit Resistance, MΩ | 0, 1 - 9999 (0=OFF) | 1 | ±(2% of setting+2counts)at 500-1000VDC, 1 -1000 MΩ |
Min-Limit Resistance, MΩ | 1 - 9999 | 1 | ±(5% of setting+2counts)at 500-1000VDC, 1000 -9999 MΩ |
±(8% of setting+2counts)at 30-500VDC, 1 -1000 MΩ |
Ramp Up Time, second | 0.1 - 999.9 | 0.1 | ±(0.1% + 0.05sec) |
Ramp Down Time, second | 0, 1.0 - 999.9 | 0.1 |
Delay Time, second | 0, 0.5 - 999.9 (0=continuous) | 0.1 |
MEASUREMENT |
AC/DC Voltage, KV | 0 - 6.00 | 0.01 | ±(1.5% of reading + 1count) |
DC Voltage, V (IR only) | 30 - 1000 | 1 | ±(1.5% of reading + 5V) |
AC Current, mA | 0 - 3.500 | 0.001 | ±(2% of reading + 2counts) |
3.00 - 20.00 | 0.01 |
DC Current, μA | 0 - 350.0 | 0.1 | ±(2% of reading + 2counts) |
DC Current, mA | 0.300 - 3.500 | 0.001 |
3.00 - 7.50 | 0.01 |
Resistance, , MΩ | 1 - 9999 (AutoRange) | 0.001 0.01 0.1 1 | ±(2% of setting+2counts)at 500-1000VDC, 1 -1000 MΩ ±(5% of setting+2counts)at 500-1000VDC, 1000 -9999 MΩ ±(8% of setting+2counts)at 30-500VDC, 1 -1000 MΩ |
GENERAL |
Input Voltage AC | 115/230Vac±15%, 50/60Hz ± 5%, Fuse 3.15A |
PLC Remote Control | Input : Test , Reset , Memory1.2.3 , Interlock |
Output : Pass , Fail , Processing , Reset-Out |
Interface(Option) | RS485 Interface |
Memory | 10 memories, 3 steps/memory |
Graphic Display | 128×64 Graphic LCD |
Safety | Built-in Smart GFI circuit,GFI trip current 450?A max, HV shut down speed: <1mS |
Key Lock | To prevent unauthorized alteration of the test Parameters |
Calibration | Build-in software and external calibrated meters |
Verification | Build-in software driven verification menu to test fault detection circuits |
Alarm Volume Setting | Range: 0-9 ;0=OFF, 1 is softest volume, 9 is loudest volume. |
Environment | 0-40℃, 20-80%RH |
Dimension/Net Weight | 215mm(W)×89mm(H)×370mm(D)/ 10Kg |
STANDARD ACCESSORIES |
Power Cord(10A) | ×1 |
Fuses | ×2 (Including a spare contained in the fuse holder) |
Interlock Disable Key(1505) | ×1 |
Hipot Test Lead ,1.5m(1101) | ×1 |
Hipot Return Lead ,1.8m(1102) | ×1 |
*product specifications are subject to change without notice. |